Dominique Reymann 2003 Metrologia 40 01005 doi:10.1088/0026-1394/40/1A/01005
Dominique Reymann
Show affiliationsIn the period 1998-2001, EUROMET organized a comparison of measurements of 10 V DC electronic reference standards. This comparison involved 22 national laboratories and the BIPM. The participating laboratories maintain their voltage reference by Josephson effect measurements or via Zener comparisons.
This paper describes the way the EUROMET.EM.BIPM-K11.b reference value xR-EUR is computed in terms of the BIPM.EM-K11.b key comparison reference value xR, i.e. how the quantity (xR-EUR - xR) and its uncertainty are estimated.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCEM according to the provisions of the Mutual Recognition Arrangement (MRA).