J H Belliss 2002 Metrologia 39 01004 doi:10.1088/0026-1394/39/1A/4
J H Belliss
Show affiliationsThis report describes a comparison of 10 pF and 100 pF capacitance standards organized within the framework of EUROMET Project No. 345 between 1995 and 1998. Piloted by the National Physical Laboratory (NPL, UK), the comparison involved sixteen national metrology institutes (NMIs), several of which are members of other regional metrology organizations, and the Bureau International des Poids et Mesures. The results show that there are significant differences between representations of the farad by some NMIs. However, the agreement demonstrated by the comparison gives confidence in maintaining traceability for the farad via either a calculable capacitor or a quantum Hall reference standard and the consensus value of the von Klitzing constant, RK-90.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database www.bipm.org/kcdb.
This Final Report has been peer-reviewed and approved for publication by the CCEM according to the provisions of the Mutual Recognition Arrangement (MRA).
06.20.Jr Determination of fundamental constants
Issue 1A (Technical Supplement 2002)
J H Belliss 2002 Metrologia 39 01004