M Boutillon 1996 Metrologia 33 479 doi:10.1088/0026-1394/33/5/7
M Boutillon
Show affiliationsAn experiment has been made in the low-energy x-ray range at the Bureau International des Poids et Mesures to estimate the change of the response of a commercial ionization chamber which, after calibration in a beam in terms of air kerma, is used in another beam. The results indicate a change in the chamber response of 0,2% to 0,5%, depending on the energy used.