Recommended values of temperature on the International Temperature Scale of 1990 for a selected set of secondary reference points

R E Bedford, G Bonnier, H Maas and F Pavese

Show affiliations


Recommended values of temperature on the International Temperature Scale of 1990 are given for a large number of secondary reference points, together with assessments of the uncertainties of these values.