A-M Perroche and M Boutillon 1995 Metrologia 32 43 doi:10.1088/0026-1394/32/1/005
A-M Perroche1,2 and M Boutillon1
Show affiliationsA critical review is presented of the parameters which are continuously or periodically verified in order to maintain the long-term stability of BIPM standards of air kerma and absorbed dose for 60Co. Some problems encountered in the measurement of the ionization current and in the determination of calibration factors are described and resolved.