R L Rusby et al 1994 Metrologia 31 149 doi:10.1088/0026-1394/31/2/009
R L Rusby1, R P Hudson2 and M Durieux3
Show affiliationsA revised table of differences (t90 - t68) between values of temperature on the International Temperature Scale of 1990, ITS-90, and the corresponding values on the International Practical Temperature Scale of 1968, IPTS-68, is published at the request of the Comité Consultatif de Thermométrie (CCT). The revision affects only the range 630°C to 1064°C, where the IPTS-68 specified the use of platinum-10% rhodium versus platinum thermocouples. It follows from new intercomparisons of thermocouples carrying IPTS-68 calibrations with platinum resistance thermometers and radiation thermometers calibrated in accordance with the ITS-90. The largest change occurs at 760°C, where the difference formerly tabulated as 0,36°C is now 0,04°C. The revision concerns only the assessment of the differences between the ITS-90 and the IPTS-68, and does not affect the ITS-90 in any way.