F Petrů et al 1992 Metrologia 29 301 doi:10.1088/0026-1394/29/4/005
F Petrů, B Popela and Z Veselá
Show affiliationsThe paper deals with He-Ne lasers at 633 nm, frequency-stabilized by saturated absorption in 127I2. The procedure of iodine cell cleaning and filling is presented in detail. This technology provides long-term reproducibility of laser frequency as may be seen from measurements taken over a four-year period. The paper shows the hyperfine spectrum of a fused silica laser with eighteen hyperfine components obtained by a third-harmonic detection technique and mean frequency differences resulting from comparisons with the standards of the PTB (Braunschweig) and the BIPM (Paris). The frequencies of the line centres of hyperfine structure (hfs) components o, p, q, r of the R (127) 11-5 absorption line of 127I2 measured directly by beat frequency measurement are presented. Taking into consideration that the measured laser has a mean frequency offset of -22 kHz for hfs components (a-n), the following frequency differences of the components o, p, q, r with reference to the component i of the reference laser INVAR 1 have been obtained: δνoi = (-263,187 ± 0,01) MHz δνqi = (-292,689 ± 0,05) MHz δνpi = (-290,271 ± 0,05) MHz δνri = (-320,720 ± 0,01) MHz. The He-Ne/I2 lasers have been used for testing the single-frequency-stabilized lasers.
42.55.Lt Gas lasers including excimer and metal-vapor lasers
42.60.Lh Efficiency, stability, gain, and other operational parameters
Issue 4 (1992)
Received 26 November 1991
F Petrů et al 1992 Metrologia 29 301
John Terblanche et al 2004 Physiol. Meas. 25 1189
Ganesh K Ramachandran et al 2005 Nanotechnology 16 1294
Wai Lam Chan et al 2007 Rep. Prog. Phys. 70 1325
Junxiang Zhang et al 2001 Europhys. Lett. 56 478
Nita Dilawar et al 2003 Metrologia 40 74
Davinder Kaur and A K Gupta 2002 J. Phys. D: Appl. Phys. 35 729
J M Gibson 2009 Rep. Prog. Phys. 72 076001
C M Surko et al 2005 J. Phys. B: At. Mol. Opt. Phys. 38 R57
S C Wayte et al 1993 Phys. Med. Biol. 38 455